| Ion Measured |
Model
Number |
Interfering
Ions***
See General Specs
for Details |
IotrodeTM
Reference
Electrode |
Lowest
+
Limit of
Detection |
Linear
Measurement
Range |
pH
Range |
Fluoride
(F-) |
AB 100 |
OH- (10) |
RC 601 |
5X10-8 Molar
.001 ppm |
Saturated - 10-6
Molar
Saturated - 0.019 ppm |
0 - 13 |
Chloride (Cl-) |
AB 110 |
OH- (80), Br-
(3X10-3),
I- (5X10-7), S2- (Trace) |
RC 401
RC 501 |
3X10-6 Molar
.105 ppm |
1 - 10-5
Molar
35,000 - 0.350 ppm |
2 - 11 |
Sulfide (S-) |
AB 120 |
Hg2+ (Trace) |
RC 501 |
10-17 Molar
3.2X10-13 ppm |
1 - 10-7
Molar
32,000 - 0.003 ppm |
11 - 13 |
Bromide (Br-) |
AB 130 |
OH- (3X104),
Cl- (400),
I- (2X10-4), S2- (Trace) |
RC 401
RC 501 |
5X10-7 Molar
.040 ppm |
1 - 10-6
Molar
80,000 - 0.080 ppm |
1 - 12 |
Iodide (I-) |
AB 140 |
OH- (108),
Cl- (106),
Br- (5X103), S2- (Trace) |
RC 401
RC 501 |
10-8 Molar
.001 ppm |
1 - 5X10-8
Molar
127,000 - 0.006 ppm |
1 - 13 |
Thiocyanate (SCN-) |
AB 150 |
OH- (3X104),
Cl- (400),
I- (2X10-4), S2- (Trace) |
RC 501 |
5X10-7 Molar
.029 ppm |
1 - 5X10-6
Molar
58,000 - 0.290 ppm |
2 - 10 |
Cyanide (CN-) |
AB 160 |
OH- (108),
Cl- (106),
Br- (5X103), S2- (Trace) |
RC 501 |
10-7 Molar
.002 ppm |
10-2 - 5X10-7
Molar
260 - 0.013 ppm |
11 - 13 |
Silver (Ag+) |
AB 170 |
Hg2+ (Trace) |
RC 601 |
10-17 Molar
1.1X10-12 ppm |
1 - 10-7
Molar
107,900 - 0.011 ppm |
1 - 13 |
Guide
Notations for 12mm Ion Selective Electrodes: ***Interfering Ions are expressed
in ratios of permissible excess. The value defines the point at which the response
to the interfering ion is more than one standard deviation of the measured ion. +
This
ratio is defined by the following quotient:
(Concentration of Interfering Ion in Molarity / Concentration of Measured Ion in Molarity) The lowest limit of detection and linear measurement range are
expressed in units of Molarity followed by the equivalents in units of parts per million
(ppm)
|