Search
Printer Friendly Version

IotrodeTM Organic Membrane Ion Selective Electrode Selection Guide

Ion Measured Model
Number
Interfering Ions***
See General Specs
for Details
IotrodeTM
Reference
Electrode
Lowest +
Limit of
Detection
Linear
Measurement
Range
pH
Range
Potassium  
(K+)
AB 400 Na+ (4X103), Li+ (4X103), NH4+ (60), Mg+2 (5X105), Ca+2 (1.5X104) RC 401
RC 501
5X10-7 Molar
.0200  ppm
1 - 5X10-6 Molar
39,000 - 0.195
2 - 12
Ammonium
(NH4+)
AB 410 Na+ (5X103), K+ (5),
H+ (1X105)
RC 401
RC 501
2X10-7 Molar
.004  ppm
0.1 - 5X10-6 Molar
1,800 - 0.090  ppm
2 - 10
Amines (Salt)
(R-NH3+ Cl-)
AB 420 Inquire to Factory RC 401
RC 501
10-5 Molar
as n-hexyl- ammonium-
chloride
0.1 - 5X10-5 Molar
as n-hexyl- ammonium-
chloride
6 - 8
Sodium
(Na+)
AB 430 Li+ (16), K+ (32), Mg+2 (2500),
Ca+2 (800)
RC 401
RC 501
10-6 Molar
.023  ppm
1 - 10-5 Molar
23,000 - 0.230  ppm
2.5 - 11
Calcium
(Ca+2)
AB 440 Na+ (102), K+ (167), Mg+2 (5X103) RC 401
RC 501
5X10-7 Molar
.020  ppm
1 - 5X10-6 Molar
40,000 - 0.200  ppm
2.5 - 11
Magnesium

(Mg+2)

AB 450 Na+ (3X104), K+ (2X103), Ca+2 (20) RC 401
RC 501
5X10-7 Molar
.012  ppm
1 - 5X10-6 Molar
24,300 - 0.121  ppm
2.5 - 11
Nitrate
(NO3-)
AB 470 ClO4- (10-3), ClO3- (0.2), Br- (6.7), Cl- (102),
HCO3- (10-1), H2PO4- (103), F- (103)
RC 401
RC 501
5X10-6 Molar
.310  ppm
1 - 10-5 Molar
62,000 - 0.620  ppm
2 - 12
Perchlorate
(ClO4-)
AB 480 NO3- (0.5), ClO3- (0.2), Br- (0.4), Cl- (20),
HCO3- (20), H2PO4- (20), F- (20)
RC 401
RC 501
5X10-6 Molar
.500  ppm
1 - 10-5 Molar
100,000 - 1.0  ppm
2 - 12
Lithium
(Li+)
AB 490 Na+ (250), K+ (200), Mg+2 (2X104), NH4+ (1X103) RC 401
RC 501
1X10-6 Molar
0.007  ppm
0.1 - 6X10-6 Molar
700 - 0.042  ppm
2 - 12
Barium
(Ba+2)
AB 500 H+ (30), Li+ (1.5X103), Na+ (300), Cs+ (550), K+ (300), NH4+ (1250)
Mg+2 (5X107), Ca+2 (50)
RC 401
RC 501
3X10-5 Molar
4.11  ppm
0.1 - 3X10-4 Molar
13,700 - 41.1  ppm
4 - 7
Cadmium
(Cd+2)
AB 510 K+, Na+, Cu2+, Ni2+, Pb2+, Ca2+
Co2+, Mg+2, Zn+2, Al3+, NH4+
Mn+2, Fe+2, Cs+, Li+, Ba+2
(approximately 103 for all ions listed)
RC 401
RC 501
10-5 Molar
1.12  ppm
10-2 - 5X10-5 Molar
1,120 - 5.6  ppm
3 - 6
Lead
(Pb+2)
AB 530 Hg2+ (0.25), Ag+ (0.32), Cu2+ (2X103), Na+ (3.1X103) Li+ (3.1X103), Cd2+ (5X103)
NH4+ (104)
RC 401
RC 501
3X10-7 Molar
0.062  ppm
1.6X10-2 - 10-6 Molar
3,315 - 0.207  ppm
3 - 6
Silver
(Ag+)
AB 540 Hg2+ (200), K+ (6.3X104), Na+, Cu2+, Ni2+, Pb2+, Ca2+ (8X104) Sr2+ (105), Co2+ (4X105) Mg+2 (2X105), Zn+2 (4X105)
Cd2+ (2.5X105)
RC 601 3X10-7 Molar
0.033  ppm
0.1 - 10-6 Molar
10,790 - 0.108  ppm
2.5 - 8.5
in AgNO3
Cesium
(Cs+)
AB 560 H+ (500), Li+ (2.5X104), Na+ (5000), Rb+ (30), K+ (300), NH4+ (600) Mg+2 (105), Ca+2 (105) RC 401
RC 501
10-5 Molar
1.33  ppm
0.1 - 3X10-5 Molar
13,290 - 3.99  ppm
2 - 12
Carbonate
(CO-3)
AB 600 SCN- (0.4), Cl- (105), Br- (6.6X103), Salicylate (.0005), HPO4- (8X104), NO3- (40) SO4- (1.25X105) RC 401
RC 501

Application
Dependent
Inquire to Factory

Application
Dependent
Inquire to Factory

Application
Dependent
Chloride
(Cl-)
AB 610 Br- (1), SCN- (1.5), Salicylate (2.5), HCO3- (3.1X105), HPO4- (8X105), SO4- (2X106),
F- (4X105), Acetate (1.25X105),
NO3- (1.5X103), ClO3- (65)
RC 401
RC 501
10-5 Molar
.355  ppm
10-1 - 5X10-5 Molar
3,545 - 1.772  ppm
2 - 12
Nitrite
(NO2-)
AB 620 SCN- (1.5X103), Benzoate (1X104),
Acetate (1X104), Salicylate (250),
HCO3- (1X104), Cl- (1X104),
NO3- (1X104)
RC 401
RC 501
5X10-6 Molar
.230  ppm
10-1 - 10-5 Molar
4,600 - .460  ppm
2 - 10
Hydronium
(H3O+ - pH)
AB 650 Na+ (1010), K+ (1010), Ca+2 (1010)
Excellent Selectivity in biological media
RC 401
RC 501
N/A N/A 6 - 8
Tetrafluoro Borate
(BF4+)
AB 660 ClO4- (0.5), SCN- (2.2), I- (2.3), ClO3- (7.6),
NO3- (40), Br- (120), CN- (125), NO2- (130),
Cl- (650), CO3- (11,500), HPO4-  (21,000), SO4- (22,000)
RC 401
RC 501
2.2X10-6 Molar
.191  ppm
10-1 - 10-5 Molar
8,680 - .868  ppm
3 - 11.5

Guide Notations for 12mm Ion Selective Electrodes:
***Interfering Ions are expressed in ratios of permissible excess.   The value defines the point at which the response to the interfering ion is more than one standard deviation of the measured ion. 
+ This ratio is defined by the following quotient: (Concentration of Interfering Ion in Molarity / Concentration of Measured Ion in Molarity)
The lowest limit of detection and  linear measurement range are expressed in units of Molarity followed by the equivalents in units of parts per million (ppm)

Advanced Sensor Technologies, Inc.
603 North Poplar Street, Orange, California 92868-1011 USA
Tel: + (1) 714-978-2837  Toll-Free: + (1) 888-969-2784 Fax: + (1) 714-978-6339 Website: http://www.astisensor.com
Manufacturing Long-Lasting Industrial pH Sensors, ORP sensors, Ion Selective Sensors & Conductivity Sensors for Difficult Inline Process Control Measurement and Monitoring Applications
Supplying Research Grade pH Electrodes, ORP Electrodes, Ion Selective Electrodes and Conductivity Electrodes for Laboratory Testing
Offering industry leading product developments and research services for OEM manufacturers and resellers requiring pH, ORP and Ion Selective Measurements for Medical In-Vitro and In-Vivo Applications and Industrial Inline, Immersion and Submersible Measurements