12mm Laboratory Organic &
Solid State Membrane Ion
Selective Electrodes
|
General Specifications for 12mm Organic
Ion Selective Electrodes:
- Electrodes are manufactured with 1 meter of cable and BNC Connector.
No other options are available at this time. See above drawing for dimension
details of ISE electrodes.
- Organic Membrane Electrodes are manufactured with rigid PVC bodies.
- Electrodes are manufactured with 16mm caps.
- Organic Membrane Electrodes are rated to 5 - 40o Celsius
Continuous, Maximum 0 - 60o Celsius Intermittent
- Standard Shelf Life is 1 year from date of manufacture.
- ***Interfering Ions are expressed
in ratios of permissible excess. The value defines the point at which the response
to the interfering ion is more than one standard deviation of the measured ion. This
ratio is defined by the following quotient:
(Concentration of Interfering Ion in Molarity / Concentration of Measured Ion in Molarity)
- The lowest limit of detection and linear measurement range are
expressed in units of Molarity followed by the equivalents in units of parts per million
(ppm)
|
General Specifications for 12mm Solid
State Ion Selective Electrodes:
- Electrodes are manufactured with 1 meter of cable and BNC Connector.
No other options are available at this time. See above drawing for dimension
details of ISE electrodes.
- Solid State Membrane Electrodes are manufactured with Ryton (PPS)
bodies.
- Electrodes are manufactured with 16mm caps.
- Solid State Membrane electrodes are rated to 5 - 50o
Celsius Continuous, Maximum -5 - 100o Celsius Intermittent
- Standard Shelf Life is 1 year from date of manufacture.
- ***Interfering Ions are expressed
in ratios of permissible excess. The value defines the point at which the response
to the interfering ion is more than one standard deviation of the measured ion. This
ratio is defined by the following quotient:
(Concentration of Interfering Ion in Molarity / Concentration of Measured Ion in Molarity)
- The lowest limit of detection and linear measurement range are
expressed in units of Molarity followed by the equivalents in units of parts per million
(ppm)
|
12mm Laboratory Reference
Electrodes
|
General Specifications for 12mm Reference Electrodes:
- Electrodes come complete with 1 meter of cable and standard Reference
Pin Connector. No other options are available at this time. See above drawing
for dimension details of reference electrodes.
- Glass construction reference electrode are manufactured with Pyrex Glass
Bodies and Large Surface Porous Ceramic Junctions. Plastic construction reference
electrode are manufactured with PVC bodies and large surface Porous Ceramic Junctions.
- Reference Model RC 301 is only recommended for applications where ease
of use and convenience (non-refillable immobilized gel system) are the primary requirement
of the application. For maximum accuracy, reference model RC 401, 501 & 601
should be employed.
- The standard pH range for All Reference Electrodes is 0-14.
- Reference Electrodes use Ag/AgCl Half-Cells.
- Reference Electrodes come standard with 16mm Caps.
- Standard Shelf Life is 1 year from date of manufacture.
- Make sure to review the recommended reference electrode for the ion
selective measurement that will be performed.
|
|
Advanced Sensor Technologies, Inc.
603 North Poplar Street, Orange, California 92868-1011 USA
Tel: + (1) 714-978-2837 Toll-Free: + (1) 888-969-2784 Fax: + (1) 714-978-6339 Website: http://www.astisensor.com
Manufacturing Long-Lasting Industrial pH Sensors, ORP sensors, Ion Selective Sensors & Conductivity Sensors for Difficult Inline Process Control Measurement
and Monitoring Applications
Supplying Research Grade pH Electrodes, ORP Electrodes, Ion Selective Electrodes and Conductivity Electrodes for Laboratory Testing
Offering industry leading product developments and research services for OEM manufacturers and resellers requiring pH, ORP and Ion Selective Measurements for Medical In-Vitro and In-Vivo Applications and Industrial Inline, Immersion and Submersible Measurements